Source: - Scott-Brown's Otorhinolaryngology, 8th Edition — Volume 2: Chapter 8: Hearing screening and surveillance, Chapter 9: Hearing tests in children, Chapter 47: Anatomy of the cochlea and vestibular system, Chapter 69: Auditory neuropathy spectrum disorder and retrocochlear disorders - Scott-Brown's Otorhinolaryngology, 9th Edition — Section 6 (Paediatrics), Chapter 125: Hearing screening and hearing tests in children (pp. 1341-1351) · Section 1 (Otology), Chapter 8: Auditory neuropathy spectrum disorder and retrocochlear disorders in adults and children (pp. 88-106) · Chapter 4: Prevention of hearing loss (CMV) · Chapter 126: Non-surgical management of the hearing-impaired child
2026
2025a) What are cochlear microphonics? b) Discuss the role of otoacoustic emissions in neonatal hearing screening. c) What is high risk neonatal hearing screening and universal neonatal hearing screening and why the later is needed to be included in NPPCD?
2023What is Universal Neonatal Hearing Screening Program? Describe the early detection and diagnosis of infant hearing impairment.
2023What is the importance of pure tone audiometry in evaluating hearing? Discuss various audiological tests for diagnosing a suspected malingerer.
2021a) What are Otoacoustic Emissions and enumerate the types of OAEs? b) Enumerate the differences between Universal Neonatal Hearing Screening and Targeted Hearing Screening.
2018a) What are different types of hearing aids? b) What are the pre-requisites of an audiometry room? (5+5)
2017a) Draw a labeled diagram of Organ of Corti. b) Function of outer hair cells. c) Otoacoustic emissions.
2014What are oto-acoustic emissions (OAE)? Briefly describe the types of OAE and their clinical applications with special reference to universal neonatal screening for hearing loss.
Discuss otoacoustic emissions including types, cochlear microphonics, their role in neonatal hearing screening, and compare universal versus targeted screening approaches. Explain the 1-3-6 rule and audiological assessment of newborns. What are Otoacoustic Emissions and enumerate the types of OAEs? Enumerate the differences between Universal Neonatal Hearing Screening and Targeted Hearing Screening. Audiological testing of newborn child; 1-3-6 rule. Cochlear microphonics; Role of OAE in neonatal hearing screening; High risk vs universal neonatal hearing screening; NPPCD inclusion.
Introduction
Neonatal hearing screening represents one of the most significant advances in paediatric audiology, enabling early identification and intervention for permanent childhood hearing impairment (PCHI). At least 1 in 1000 children is born with permanent bilateral hearing impairment of 40 dB HL or greater. The rationale for early detection is compelling: if PCHI is identified before 6 months of age and habilitation started soon thereafter, the adverse effects on speech, language, and cognitive development are significantly lessened.
The two primary screening technologies are otoacoustic emissions (OAEs) and automated auditory brainstem response (AABR). Understanding the physiological basis of these tests, particularly OAEs and cochlear microphonics, is essential for effective screening programme implementation.
Otoacoustic Emissions (Oaes)
Definition and Physiological Basis
Otoacoustic emissions are low-intensity sounds generated by the active movement of the outer hair cells (OHCs) of the cochlea. They represent acoustic energy that is emitted from the cochlea back through the middle ear into the external auditory canal, where they can be recorded using sensitive microphone systems.
The physiological mechanism underlying OAEs involves the cochlear amplification process:
Sound-induced travelling waves along the basilar membrane stimulate outer hair cells
OHCs exhibit unique fast reversible axial length changes (electromotility) driven by the motor protein prestin
This active process amplifies and fine-tunes the signal before it reaches the inner hair cells
The excess energy released by this cochlear amplification generates sound waves emitted as OAEs
This phenomenon was predicted by Thomas Gold in 1948 and demonstrated by David Kemp in 1978. OAEs provide an objective measure of the "health" of the outer hair cells and can be recorded from probe microphones fitted in the external ear canal.
Characteristics of OAEs
Present in ears with normal cochlear function
Absent if there is damage to the outer hair cells
Can be affected by middle ear pathology (fluid/debris)
Do not require neural pathway integrity
Provide frequency-specific information about cochlear function
Non-invasive and quick to perform
Types of Otoacoustic Emissions
1. Spontaneous Otoacoustic Emissions (SOAEs)
Definition: Low-level sounds emitted from the cochlea in the absence of any external acoustic stimulation.
Characteristics:
Present in approximately 40-60% of normal-hearing individuals
More common in females than males
More common in right ears than left ears
Typically narrow-band signals at specific frequencies
Clinical utility is limited due to variable presence
Definition: Emissions generated when two pure tones of different frequencies (f1 and f2) are presented simultaneously.
Mechanism:
Non-linear interaction of the two primary tones in the cochlea
Most prominent distortion product is at frequency 2f1-f2
Optimal f2/f1 ratio is approximately 1.22
Clinical features:
Frequency-specific assessment possible
Can test wider frequency range (up to 8 kHz)
Better high-frequency assessment than TEOAEs
Useful for monitoring ototoxicity
Takes longer than TEOAE testing
4. Stimulus Frequency Otoacoustic Emissions (SFOAEs)
Definition: Emissions evoked at the same frequency as a continuous pure-tone stimulus.
Characteristics:
Technically more difficult to record
Requires separation of emission from stimulus
Primarily used in research settings
Limited clinical application currently
Comparison of OAE Types
TEOAEs:
Broadband click stimulus
Quick screening tool
Frequency range 1-4 kHz
Best for screening purposes
DPOAEs:
Two-tone stimulus
Frequency-specific
Extended frequency range
Better for diagnostic and monitoring purposes
9e (Ch 125): OAE uses are screening, monitoring cochlear damage (e.g. from ototoxic agents or noise), separating sensory from neural dysfunction (ANSD, auditory processing disorder, acoustic neuroma) and objective evidence of cochlear function in non-organic loss or difficult-to-test patients. Normal PTA can coexist with abnormal OAEs, and mild or frequency-specific loss with normal OAEs.
Cochlear Microphonics
Definition
Cochlear microphonics (CM) are electrical potentials generated in the cochlea that mimic the acoustic properties of the stimulus. They are an electrophysiological response generated primarily from the outer hair cells to acoustic signals.
Characteristics
Start at near-zero latency (appear immediately after stimulus)
Represent early components of cochlear response
Generated primarily by outer hair cells
Invert in polarity when stimulus phase is reversed (rarefaction to condensation)
No latency shift with intensity decrease
Small in amplitude and susceptible to electroacoustic artefact
Recording Technique
Cochlear microphonics are incorporated into auditory brainstem response (ABR) testing software. To evaluate for CM:
ABR is evoked using rarefaction and condensation clicks at suprathreshold levels
CM appears as peaks that invert in amplitude when stimulus polarity is switched
Observation of this inversion is consistent with detection of CM
Clinical Significance
In Auditory Neuropathy Spectrum Disorder (ANSD):
CM testing is critical for diagnosing ANSD, which is characterized by:
Normal to raised pure-tone thresholds
Poor speech discrimination
Delayed or absent ABRs
Normal OAEs and cochlear microphonics
The presence of CM with absent ABR indicates:
Normal preneural (outer hair cell) function
Disruption at inner hair cells, synapses, or auditory nerve
9e (Ch 8: ANSD and retrocochlear disorders):
Otoferlin (OTOF) mutations are associated with non-syndromic recessive auditory neuropathy (NSRAN).
NICU-treated children have a higher incidence of AN; the 2019 JCIH update recommends ABR testing for NICU infants, because AN may be missed by an OAE-based universal screen. Perinatal factors include prematurity, anoxia, low birth weight, hyperbilirubinaemia, ototoxicity and ventilator dependency.
9e (Ch 125): in the NICU pathway AOAE can be present with AABR absent, so immediate referral for diagnostic ABR is needed to confirm ANSD.
Pitfalls in CM Recording
Stimulus artefact confusion:
Artefact arises from the transducer, not the cochlea
Shows similar morphology to CM
Also inverts with polarity change
Risk increases at stimulus levels above 80 dB nHL
Tube insert earphones with longer tubing can separate artefact from CM in time
Comparison: CM, Compound Action Potential, and OAE
Feature
Cochlear Microphonic
Compound Action Potential
OAE
Source
Primarily OHCs
Auditory nerve
OHCs
Measurement
Electrical recording
Electrical recording
Acoustic recording
Polarity inversion
Yes, follows stimulus
No
N/A
Latency shift
No
Yes, with intensity
N/A
Neonatal Hearing Screening
Rationale for Newborn Screening
Prevalence:
At least 1 in 1000 children born with bilateral PCHI ≥40 dB HL
Prevalence increases through childhood (progressive/acquired losses)
Consequences of late identification:
Compromised communication skills
Literacy difficulties
Behavioural problems
Reduced educational achievement
Mental health issues
Impaired family dynamics
Reduced quality of life
Benefits of early identification:
Confirmation of PCHI by 9 months associated with better receptive language
HVDT phased out. Before UNHS the UK used the Health Visitor Distraction Test (HVDT) as a universal 8-month screen (8e); it was phased out for poor sensitivity, poor specificity, high refer rate, low yield and poor cost per case found. 9e (Ch 125): HVDT was done at 6-9 months and detected more temporary than permanent losses; the 1997 HTA critical review recommended universal screening for moderate or worse PCHI in the better ear.
Evidence for UNHS (8e Ch 8). The only quasi-RCT is the Wessex trial: birth during UNHS periods was associated with confirmation of PCHI by 9 months and better receptive language (not speech) at age 8. The 2008 UNHS Task Force and two later population studies (outcomes at 3-5 and 7-8 years) support it. Outcomes at 3 years were only weakly related to age at hearing-aid fitting, but for cochlear implant users the age at switch-on was significantly associated with better outcomes.
Current Screening Tests
1. Automated Otoacoustic Emissions (AOAE)
Technology:
Portable hand-held screening instruments
Automated detection algorithms
Uses transient or distortion product emissions
No staff judgement required for pass/refer decision
Advantages:
Quick (typically 2-5 minutes)
Minimally invasive
Low consumable costs
High sensitivity for cochlear dysfunction
Disadvantages:
Affected by outer/middle ear debris or fluid
Will miss auditory neuropathy spectrum disorder
Cannot assess neural pathway integrity
2. Automated Auditory Brainstem Response (AABR)
Technology:
Measures neural response from brainstem
Automated detection algorithms
Electrodes placed on baby's head
Advantages:
Assesses entire auditory pathway including neural component
Definition: Screening all newborn babies regardless of risk factor status.
Advantages:
Identifies ALL congenital cases (not just 60%)
Better coverage rates (>98%)
Earlier median age of identification (~7 weeks)
Lower loss to follow-up rates
Higher positive predictive value
Cost-effective in long-term
Parental preference (80-90% want it)
Avoids parental anxiety from delayed detection
Performance metrics (English NHSP):
Coverage: 98.1% by 4 weeks corrected age
Refer rate: 2.5% (0.5% bilateral)
Yield: 1.01/1000 for bilateral PCHI
Median age of identification: ~7 weeks
Comparison Table
Parameter
Universal Screening
Targeted Screening
Population screened
All babies (100%)
High-risk only (~10%)
Cases identified
~100%
~50-60%
Coverage achievable
>98%
Variable
Cost per case found
Lower long-term
Higher
Loss to follow-up
Lower
Higher
Parental acceptance
High (>99%)
N/A
Mild PCHI evidence is equivocal (8e Ch 8). Wake et al. population study of children with slight/mild sensorineural hearing loss found no strong evidence for adverse effects on language, reading, behaviour or health-related quality of life in children who are otherwise healthy and of normal intelligence. The optimal form of systematic intervention for mild PCHI (including the role and benefits of amplification) remains to be determined. This is why most national UNHS programmes target moderate-or-worse PCHI (≥ 40 dB) rather than any detectable loss.
The 1-3-6 Rule
Definition
The 1-3-6 rule represents the recommended timeline benchmarks for early hearing detection and intervention (EHDI) programmes:
1 Month: Hearing screening completed
3 Months: Audiological diagnosis confirmed
6 Months: Intervention initiated (hearing aids fitted, early intervention started)
Rationale
Research evidence demonstrates that identification of PCHI before 6 months of age and early intervention leads to:
Better receptive language scores at age 8 years
Better speech outcomes
Improved educational achievement
Better psychosocial development
Components of the Timeline
By 1 Month - Screening:
All babies screened before hospital discharge or within first month
Rescreening of initial refers completed
Referral to audiology for screen-positive cases
By 3 Months - Diagnosis:
Comprehensive audiological evaluation completed
Type, degree, and configuration of hearing loss determined
87% of screen referrals enter follow-up by 4 weeks
95% enter follow-up by 6 months
Median age of identification: approximately 7 weeks
The 1-3-6 timeline itself comes from the US JCIH guidance; Scott-Brown does not state it. 9e (Ch 125) cites the 2007 JCIH statement (an update of 2000): early detection and intervention to maximise language and literacy, identification of all types of hearing loss, ongoing monitoring and medical evaluation for the cause.
Audiological Assessment of Newborns
Electrophysiological Tests
1. Auditory Brainstem Response (ABR)
Purpose: Threshold estimation and neural pathway assessment
Technique:
Click or tone-burst stimuli
Electrodes on forehead and mastoids
Can be performed during natural sleep in first 6 months
Applications:
Threshold estimation
ANSD diagnosis (with CM evaluation)
Neural pathway integrity assessment
2. Auditory Steady-State Response (ASSR)
Advantages:
Frequency-specific thresholds
Can test at higher intensities than ABR
Useful for severe-profound losses
Limitations:
Affected by state of arousal
May be problematic in difficult-to-test populations
Objective Cochlear Tests
1. Otoacoustic Emissions
As described above
Confirms OHC function
Combined with ABR for comprehensive assessment
2. Acoustic Reflexes
Middle ear muscle reflex testing
Absent in conductive loss and ANSD
Useful adjunct to OAE and ABR
Behavioural Tests
Behavioural Observation Audiometry (BOA)
Age range: 0-6 months
Technique:
Observe behavioural responses to calibrated sounds
Eye widening, startle, arousal changes
Cessation or changes in activity
Limitations:
Subjective interpretation
Cannot determine thresholds accurately
High false-negative rate
Visual Reinforcement Audiometry (VRA)
Age range: 6-30 months (can sit with support, typically 5-7 months onwards)
Technique:
Sound presented, head turn reinforced by visual reward
Can obtain ear-specific and frequency-specific information
More reliable than BOA
Assessment Protocol for Screen Referrals
History: Risk factors, developmental milestones, family history
Otoscopy: Middle ear status
Tympanometry: Middle ear function (age-appropriate probe tone)
OAE: Cochlear (OHC) function
ABR with CM: Neural pathway and preneural function
Behavioural testing: Age-appropriate (BOA, VRA)
🆕 Part G (9e): Screen Exclusions and Surveillance After the Screen
🆕 9e — NHSP England pathways (Ch 125)
NHSP was introduced in 2002 and universal across England by March 2006; coverage was 99.2% in 2016/2017. PCHI of moderate or worse degree is 1-2 per 1000 births, rising to about 1 in 100 after more than 48 hours in intensive care.
Well baby: AOAE1, then AOAE2 if no clear response, then AABR; a pass at any stage is discharge. NICU: both AOAE and AABR. An AABR fail in one or both ears means referral within 4 weeks of screen completion (or by 44 weeks gestational age) for diagnostic ABR. An AOAE fail in both ears, or risk factors, means targeted assessment at 7-9 months by VRA; an AOAE fail in one ear with no risk factors is discharged.
Four screen exclusions go straight to diagnostic ABR: microtia and canal atresia; neonatal bacterial meningitis or meningococcal septicaemia (urgent, because labyrinthine ossification may need urgent cochlear implantation); programmable VP shunt (BSA 2023); confirmed congenital CMV (short window for antivirals).
🆕 9e — After a passed screen (Ch 125)
A retrospective NHSP cohort of nearly 4 million births over 6 years found the evidence for routine targeted surveillance of babies who pass is not strong; it is not recommended.
Late presentation: mild or frequency-specific losses are missed by NHSP; congenital CMV and EVA can present later; acquired causes (meningitis, head injury, OME) follow. PCHI over 40 dB HL is 1.07 per 1000 at 3 years and 2.05 per 1000 by 9-15 years.
School entry screening (since 1955) finds fewer cases since NHSP; a 2007 review found wide variation and just over 10% of services had abandoned it; some UK areas still provide it.
8e (Ch 8): surveillance is needed because of acquired, late-onset and progressive loss, missed cases, mild and high-frequency loss and children moving from countries without screening. Risk-factor surveillance after a pass was then kept only for: syndrome (other than Down) linked to hearing loss, NICU with OAE refer in both ears and AABR pass in both, craniofacial anomaly, Down syndrome and congenital infection. Assess after meningitis, ECMO, skull fracture and hyperbilirubinaemia. Services must respond to any parental or professional concern with age-appropriate assessment. Evidence for the school-entry screen is equivocal.
🆕 9e — Congenital CMV (Ch 4, Ch 126)
Diagnosis needs virus in urine or saliva within the first 3 weeks (saliva PCR preferred). SNHL develops in 30-40% of symptomatic and 10-15% of asymptomatic cases (late onset at an average of 44 months); progression 50%, fluctuation 23%, so long-term monitoring is needed.
Antivirals started within the first month (6 weeks IV ganciclovir or 6 months oral valganciclovir) improve audiological and neurodevelopmental outcome. After 3 months, a positive IgG leads to CMV DNA PCR on the Guthrie dried blood spot (Ch 126).
END OF ANSWER
Based on Scott-Brown's Otorhinolaryngology, 8th & 9th Editions